METHOD DEVELOPMENT, VALIDATION AND STABILITY STUDIE S FOR DETERMINATION OF LURASIDONE HYDROCHLORIDE IN TABLET DOSAGE FORM BY R P-HPLC
By: Thota, Bhavani.
Contributor(s): Adepu, Geetha Susmita.
Publisher: M P ABI Prints & Publishing Co. 2018Edition: Vol.10(12).Description: 58-63p.Subject(s): PHARMACEUTICSOnline resources: Click here In: International journal of pharmacy and pharmaceutical scienceSummary: Objective: To develop a simple, rapid, sensitive, precise, acc urate, economical and validated reverse phase high performance liquid chromatographic (RP-HPLC) method for the estimation of lurasidone hydrochloride in tablet dosage form. Methods: The chromatographic separation was carried out on a prontosil C18, AQ (100 mm×4.6 mm, 3 μm) column. A mixture of phosphate buffer (pH 3.0): acetonitrile (ACN) (55:45v/v) was used as a mobile phase. Flow rate of 1.0 ml/min and 10 μl i njection volume was used for the assay. PDA detector was used, and the detection wavelength was 230 nm. The retention time (RT) of lurasidone hydr ochloride was found to be 4.505±0.01 min. The method was validated according to the ICH guide lines. Results : The calibration curve for lurasidone hydrochloride was linear with a correlation coefficient value 0.9 99 in the concentration range of 25- 125%. Specificity, accuracy (% mean recovery, 99.08 %), precision, detection limits, robustness (% RSD˂ 2) and system suitability were found to be within limits. Degradation studies were performed u nder different stressed conditions, and the results of degradation studies reveal that the developed method was stable. Conclusion: The developed method was simple, reliable, economic al and stable and it can be applied for the routine quality control analysis of lurasidone hydrochloride in tablet dosage forms.Item type | Current location | Call number | Status | Date due | Barcode | Item holds |
---|---|---|---|---|---|---|
Articles Abstract Database | School of Pharmacy Archieval Section | Not for loan | 2020957 |
Objective:
To develop a simple, rapid, sensitive, precise, acc
urate, economical and validated reverse phase high
performance liquid
chromatographic (RP-HPLC) method for the estimation
of lurasidone hydrochloride in tablet dosage form.
Methods:
The chromatographic separation was carried out on a
prontosil C18, AQ (100 mm×4.6 mm, 3 μm) column. A
mixture of phosphate buffer
(pH 3.0): acetonitrile (ACN) (55:45v/v) was used as
a mobile phase. Flow rate of 1.0 ml/min and 10 μl i
njection volume was used for the assay. PDA
detector was used, and the detection wavelength was
230 nm. The retention time (RT) of lurasidone hydr
ochloride was found to be 4.505±0.01 min.
The method was validated according to the ICH guide
lines.
Results
:
The calibration curve for lurasidone hydrochloride
was linear with a correlation coefficient value 0.9
99 in the concentration range of 25-
125%. Specificity, accuracy (% mean recovery, 99.08
%), precision, detection limits, robustness (% RSD˂
2) and system suitability were found to be
within limits. Degradation studies were performed u
nder different stressed conditions, and the results
of degradation studies reveal that the
developed method was stable.
Conclusion:
The developed method was simple, reliable, economic
al and stable and it can be applied for the routine
quality control analysis of
lurasidone hydrochloride in tablet dosage forms.
There are no comments for this item.